Sper Xjentifiku 860033

Manwal tal-Utent tal-Miter tal-Kwalità tal-Ilma tal-Benchtop Sper Scientific Mudell 860033

Multi-Parameter pH, ORP, Conductivity, TDS, Salinity & Temperature Analyzer

1. Introduzzjoni

The Sper Scientific Benchtop Water Quality Meter Model 860033 is a multi-parameter instrument designed for precise measurement of various water quality indicators. This device accurately measures pH, Oxidation-Reduction Potential (ORP), millivolts (mV), conductivity, Total Dissolved Solids (TDS), salinity, and temperature. It is suitable for a wide range of applications in laboratories and industrial environments, offering features such as automatic calibration and temperature compensation.

Sper Scientific Benchtop Water Quality Meter 860033 displaying pH reading

Figure 1: Sper Scientific Benchtop Water Quality Meter Model 860033 in operation, measuring pH in a beaker.

2. Setup

2.1 Żpakkjar u Komponenti

Iftaħ il-komponenti kollha bir-reqqa mill-pakkett u vverifika li l-oġġetti li ġejjin huma inklużi:

  • Sper Scientific Benchtop Water Quality Meter 860033
  • Conductivity Electrode (also measures TDS and Salinity)
  • pH Electrode (also measures mV)
  • Multiple Electrodes Holder Arm
  • Adapter AC
  • Kejbil tal-Kompjuter
  • Software CD
All components of the Sper Scientific Benchtop Water Quality Meter 860033

Figure 2: All included components of the Benchtop Water Quality Meter, including the main unit, electrodes, power adapter, and accessories.

2.2 Connecting Electrodes and Power

  1. Attach the Electrode Holder Arm: Securely attach the multiple electrodes holder arm to the side of the meter unit.
  2. Qabbad l-Elettrodi: Connect the pH electrode to the BNC connector on the rear panel. Connect the conductivity electrode to its designated port.
  3. Connect AC Adaptor: Plug the AC adaptor into the DC input port on the rear of the meter, then connect it to a standard power outlet. The meter operates on AC voltage and does not use batteries.
  4. Optional PC Connection: For data logging and analysis, connect the computer cable to the RS232 port on the meter and to your computer. Install the provided software.
Rear panel connections of the Sper Scientific Benchtop Water Quality Meter 860033

Figure 3: Rear panel of the meter showing DC power input, RS232 port, analog outputs, and BNC/ATC connections for electrodes.

3. Istruzzjonijiet Operattivi

3.1 Power On/Off

Agħfas il- QAWWA button to turn the meter on. The large LCD will display the current parameter, time, date, and temperature. Press the QAWWA button again to turn the meter off.

3.2 Għażla tal-Parametru

Uża l- MODALITÀ button to cycle through the available measurement parameters: pH, ORP, mV, Conductivity, TDS, Salinity, and Temperature.

3.3 Kalibrazzjoni

The meter features automatic buffer recognition for pH calibration and supports 5 calibration points. Refer to the detailed calibration procedure in the full instruction manual for each specific parameter (pH, ORP, Conductivity, TDS, Salinity).

  • Kalibrazzjoni tal-pH: Use standard pH buffer solutions. The meter will automatically recognize the buffer values.
  • Conductivity/TDS/Salinity Calibration: Use appropriate standard solutions.

3.4 Teħid tal-Kejl

  1. Ensure the appropriate electrode is connected and properly conditioned.
  2. Select the desired measurement parameter using the MODALITÀ buttuna.
  3. Immerse the electrode(s) into the sample solution. Ensure the sensor part of the electrode is fully submerged.
  4. Allow the reading to stabilize. The meter provides automatic or manual temperature compensation and ranging.
  5. The LCD will display the stable measurement.
  6. Uża l- HOLD/REC button to hold the current reading or record data (if connected to PC with software).

4. Manutenzjoni

4.1 Kura tal-Elettrodi

  • Elettrodu tal-pH: Always keep the pH electrode bulb moist. Store it in a suitable storage solution or pH 4 buffer when not in use. Never store in distilled water.
  • Conductivity Electrode: Rinse the conductivity electrode with distilled water after each use. Store dry or in a conductivity storage solution as recommended.
  • Tindif: Clean electrodes regularly according to the full manual's instructions to prevent buildup that can affect accuracy.

4.2 Meter Cleaning

Wipe the meter's housing with a soft, damp cloth. Do not use abrasive cleaners or solvents. Ensure no liquid enters the meter's ports or openings.

4.3 Ħażna

When not in use for extended periods, disconnect the AC adaptor and store the meter and electrodes in a clean, dry environment within the specified operating temperature range.

5 Issolvi l-problemi

This section addresses common issues you might encounter with your Sper Scientific Benchtop Water Quality Meter 860033.

  • Qari Instabbli: If readings are unstable, ensure electrodes are properly conditioned and calibrated. Gently stir the sample solution around the probe to help readings settle. Check for air bubbles on the electrode surface.
  • Qari mhux eżatt: Recalibrate the meter with fresh standard solutions. Check electrodes for damage or contamination. Ensure temperature compensation is correctly applied.
  • Meter Not Powering On: Verify the AC adaptor is securely connected to both the meter and a working power outlet.
  • No Communication with PC: Ensure the computer cable is correctly connected and the software drivers are installed. Check the RS232 port settings on your computer.

For more detailed troubleshooting steps, please refer to the comprehensive instruction manual provided with your device or contact Sper Scientific customer support.

6. Speċifikazzjonijiet

The following table outlines the measurement ranges, resolutions, and accuracies for the Sper Scientific Benchtop Water Quality Meter Model 860033.

Table of specifications for Sper Scientific Benchtop Water Quality Meter 860033

Figure 4: Detailed specifications including measurement ranges, resolutions, and accuracies for pH, mV/ORP, Conductivity, TDS, Salinity, and Temperature.

Speċifikazzjonijiet Ġenerali
KaratteristikaDettall
Dimensjonijiet8.5-il pulzier (L) x 6.5-il pulzier (W) x 2.25-il pulzier (H)
Sors tal-EnerġijaAC Adaptor (no batteries required)
OutputsDigital (RS232), Analog
Kumpens tat-TemperaturaAwtomatiku jew Manwal
pH Buffer RikonoxximentAwtomatiku
Punti ta' Kalibrar5
ĊertifikazzjonijietApprovat CE

7. Informazzjoni dwar Garanzija

Sper Scientific products are manufactured under strict quality control standards. For specific warranty terms and conditions, please refer to the warranty card included with your product or visit the official Sper Scientific websit. Żomm l-irċevuta tax-xiri tiegħek bħala prova tax-xiri għal kwalunkwe talba ta' garanzija.

8. Appoġġ għall-Klijent

If you require technical assistance, have questions about your meter, or need to order replacement parts, please contact Sper Scientific customer support. Contact information can typically be found on the product packaging, the official website, or the full instruction manual.

Sper Scientific Contact Information:

  • Websit: www.sperscientific.com (for product information, FAQs, and support resources)
  • Email/Telefon: Irreferi għall- website or product documentation for the most current contact details.

Dokumenti Relatati - 860033

Preview Manwal tal-Operazzjoni tal-Analizzatur tal-Ossiġnu Maħlul li Jista' Jinġarr
Dan il-manwal tat-tħaddim jipprovdi struzzjonijiet dettaljati għall-Analizzatur tal-Ossiġnu Maħlul li Jista' Jinġarr, li jkopri informazzjoni dwar is-setup, il-kalibrazzjoni, it-tħaddim, il-manutenzjoni, u s-sigurtà għal monitoraġġ preċiż tal-kwalità tal-ilma.
Preview Manwal tal-Utent u Speċifikazzjonijiet tal-Pinna tal-Ossiġnu Maħlul Sper Scientific 850045
Manwal tal-utent u speċifikazzjonijiet għall-Pinna tal-Ossiġnu Maħlul Sper Scientific 850045. Tgħallem dwar il-karatteristiċi tagħha, kif tikkalibra u tkejjel l-ossiġnu maħlul u t-temperatura, il-manutenzjoni tas-sonda, is-sostituzzjoni tal-batterija, l-ispeċifikazzjonijiet tekniċi, u informazzjoni dwar il-garanzija.
Preview Manwal tal-Istruzzjonijiet tal-Miter UVAB/UVC tad-Datalogging Sper Scientific 850031
Manwal ta' struzzjoni għall-Mudell 850031 tal-Meter tad-Datalogging UVAB/UVC Sper Scientific, li jiddeskrivi l-karatteristiċi tiegħu, il-proċeduri ta' kejl għall-UVA, UVC, u t-temperatura, il-funzjonijiet tad-datalogging, is-settings avvanzati, l-ispeċifikazzjonijiet, u l-garanzija. Iddisinjat għal applikazzjonijiet industrijali, xjentifiċi, u ta' sigurtà.
Preview Manwal ta' Istruzzjoni tar-Refrattometru Diġitali Sper Scientific
Manwal ta' struzzjonijiet għar-Refrattometri Diġitali Sper Scientific (Mudelli 300033, 300034, 300035, 300036), li jkopri t-tħaddim, il-kalibrazzjoni, l-ispeċifikazzjonijiet, u l-garanzija.
Preview Manwal tal-Operazzjoni tad-Detector tal-Kwalità tal-Arja Sper Scientific
Dan il-manwal jipprovdi struzzjonijiet dwar kif tħaddem id-Detector tal-Kwalità tal-Arja Sper Scientific, inkluż is-setup, is-settings tas-sistema, ir-rekords tal-istorja, il-korrezzjoni tas-sensur, il-modifika tad-dejta, u l-prontijiet tal-allarm.
Preview Termometru Infrared ta' Użu Doppju mingħajr Kuntatt Sper Scientific TE-94 - Kejl Preċiż tat-Temperatura
Skopri t-Termometru Infra-aħmar b'Użu Doppju mingħajr Kuntatt Ċertifikat NIST Sper Scientific TE-94. Joffri preċiżjoni, konvenjenza, u kejl avvanzat mingħajr kuntatt għat-temperaturi tal-ġisem u tal-wiċċ. Ideali għall-użu fid-dar, fl-uffiċċju, u industrijali.