Elitech VG-760

Manwal tal-Istruzzjonijiet tal-Gauge tal-Vakwu tal-Mikron Diġitali tal-HVAC Elitech VG-760

1. Introduzzjoni

This manual provides detailed instructions for the proper use, setup, operation, maintenance, and troubleshooting of the Elitech VG-760 Digital HVAC Micron Vacuum Gauge. This device is designed for accurate vacuum measurement in HVAC and refrigeration systems, ensuring efficient and reliable performance.

The Elitech VG-760 is a robust digital vacuum gauge featuring a wide measurement range, high accuracy, and multiple measurement units. It is built with an IP65 rating for enhanced protection against environmental factors.

2. Informazzjoni dwar is-Sigurtà

  • Always wear appropriate personal protective equipment (PPE) when working with HVAC/R systems.
  • Ensure the system is depressurized before connecting or disconnecting the vacuum gauge.
  • Do not expose the gauge to extreme temperatures or corrosive substances.
  • Handle the gauge with care to prevent damage to the sensor.
  • Żomm l-apparat 'il bogħod mit-tfal.

3. Prodott Aktarview

The Elitech VG-760 Digital HVAC Micron Vacuum Gauge is designed for precise vacuum measurement. Its compact design and clear display make it suitable for various applications.

Elitech VG-760 Digital Vacuum Gauge

Figure 1: Elitech VG-760 Digital Vacuum Gauge, showing its compact design and digital display.

3.1 Karatteristiċi ewlenin

  • Wide Measurement Range: 0-19000 Microns.
  • High Accuracy: ±10% or ±10 Micron.
  • Multiple Measurement Units: inHg, Torr, psi, mbar, mTorr, Pa, micron, Kpa.
  • Settable High/Low Pressure Alarms.
  • IP65 Rated for enhanced protection.
  • Long Battery Life: Up to 300 hours with 3 AA batteries.

3.2 Funzjonijiet tad-Displej u l-Buttuni

Elitech VG-760 display and button functions

Figure 2: Elitech VG-760 display elements and button controls. 1. Battery Level, 2. Auto OFF (15 min), 3. Current Vacuum Value, 4. Units. Buttons: A (Backlight on/off), B (Power on/off), C (Switch units). Pressing A+B enables auto-off.

Elitech VG-760 display showing 8 units

Figure 3: The Elitech VG-760 supports 8 different measurement units: inHg, Torr, psia, mbar, mTorr, Pa, micron, and kPa.

3.3 Filtering Components

Elitech VG-760 multiple filtering components

Figura 4: Sploda view of the Elitech VG-760's multiple filtering system, including the O-ring, Filter Net, Filter Cotton, and 1/4" SAE Fitting, designed to protect the sensor from contaminants.

4. Setup

4.1 Disimballaġġ

Upon receiving your Elitech VG-760, carefully unpack all components. The packing list typically includes:

Elitech VG-760 packing list and contents

Figure 5: Contents of the Elitech VG-760 package, including the vacuum gauge, user manual, AA batteries, zip bag, filter cotton, filter net, and O-rings.

  • 1 x Vacuum Gauge
  • 1 x Manwal għall-Utent
  • 3 x Batteriji AA
  • 1 x Zip Bag
  • 2 x Filter Cotton & Filter Net
  • 2 x O-ring

4.2 Installazzjoni tal-batterija

  1. Locate the battery compartment on the back of the gauge.
  2. Iftaħ l-għatu tal-kompartiment.
  3. Daħħal 3 batteriji AA, filwaqt li tiżgura polarità korretta.
  4. Agħlaq l-għatu tal-kompartiment tal-batterija sew.

4.3 Konnessjoni mas-Sistema

The VG-760 features a 1/4" SAE male flare fitting for connection to HVAC/R systems. Ensure all connections are tight to prevent leaks.

Video 1: Demonstrates how to connect the Elitech VG-760 to an HVAC system using a T-connector and bend adapter for vacuum measurement. It highlights the oil-proof film protection on the sensor and the process of attaching the gauge to a vacuum pump.

  1. Attach the VG-760 to the system using appropriate fittings (e.g., T-connector and bend adapter).
  2. Ensure the connection is secure and leak-free.

5. Istruzzjonijiet Operattivi

5.1 Power On/Off

Agħfas il- Qawwa button (center button) to turn the gauge on or off.

5.2 Kontroll tad-dawl ta' wara

Agħfas il- Dawl ta' wara button (left button) to toggle the display backlight on or off for improved visibility in various lighting conditions.

5.3 Switching Measurement Units

Agħfas il- UNITÀ button (right button) to cycle through the available measurement units: inHg, Torr, psi, mbar, mTorr, Pa, micron, and Kpa.

5.4 Funzjoni Auto-Off

The gauge features an auto-off function to conserve battery life. To enable or disable this feature, press and hold both the Dawl ta' wara u Qawwa buttuni fl-istess ħin.

5.5 Interpretazzjoni tal-Qari

The gauge displays the vacuum value in real-time. If the display shows dashes (-----), it indicates that the current pressure exceeds the measurement range of the gauge (above 19,000 microns).

6. Manutenzjoni

6.1 Tindif tas-Sensor

If the vacuum gauge sensor becomes contaminated with oil, it may display no value or inaccurate readings. In such cases, it is necessary to clean the vacuum sensor.

Video 2: Illustrates the importance of cleaning the vacuum sensor if it becomes contaminated with oil, as contamination can prevent accurate readings. The video shows how to access and clean the sensor using a solvent.

  1. Disconnect the gauge from the system.
  2. Carefully remove the 1/4" SAE fitting to expose the sensor.
  3. Use a suitable solvent (e.g., isopropyl alcohol) and a lint-free cloth or cotton swab to gently clean the sensor surface.
  4. Allow the sensor to dry completely before reassembling.
  5. Replace the O-ring, filter net, and filter cotton as needed to ensure proper sealing and filtration.

6.2 Kura Ġenerali

  • Store the gauge in its protective zip bag when not in use.
  • Avoid dropping or subjecting the gauge to strong impacts.
  • Imsaħ il-barra b'artab, damp cloth as needed. Do not use abrasive cleaners.

7 Issolvi l-problemi

ProblemaKawża PossibbliSoluzzjoni
Gauge does not power onBatteriji mejta jew installati ħażinIbdel il-batteriji jew iċċekkja l-polarità
Display shows dashes (-----)Pressure is above 19,000 microns (out of range)Continue evacuation until pressure drops within range
Inaccurate or no readingContaminated sensor (oil, debris)Clean the vacuum sensor as per maintenance instructions
Inaccurate or no readingTnixxija tas-sistemaCheck all connections for leaks and repair as necessary

8. Speċifikazzjonijiet

Elitech VG-760 dimensions and technical specifications

Figure 6: Elitech VG-760 dimensions and key technical specifications.

ParametruValur
Dimensjonijiet tal-Prodott4.92 x 1.38 x 2.76 pulzieri
Piż tal-oġġett15.52 Ounces (0.6 lbs)
Numru tal-Mudell tal-OġġettVG-760
Batteriji3 batteriji AA (inklużi)
ManifatturElitech
MaterjalPlastik
Medda ta' Kejl0-19000 mikron
Eżattezza±10% or ±10 Micron
Temperatura operattiva0 ° F ~ 140 ° F (-17.8 ° C ~ 60 ° C)
Armar1/4" SAE Male Flare
Ħajja tal-batterijaSa 300 sigħat
Livell ta' ProtezzjoniIP65 Rated

9. Informazzjoni dwar Garanzija

Elitech products are manufactured to high-quality standards. For specific warranty terms and conditions, please refer to the warranty card included with your product or contact Elitech customer support. Keep your purchase receipt as proof of purchase for warranty claims.

10. Appoġġ għall-Klijent

For technical assistance, troubleshooting, or any inquiries regarding your Elitech VG-760 Digital HVAC Micron Vacuum Gauge, please contact Elitech customer support.

  • Appoġġ bl-Email: Disponibbli 24/7
  • Appoġġ tat-Telefon: Available for US customers

Please visit the official Elitech webis-sit għall-aktar informazzjoni ta' kuntatt u riżorsi ta' appoġġ aġġornati.

Dokumenti Relatati - VG-760

Preview Gauge tal-Manifold Diġitali Elitech MS-100: Karatteristiċi, Tħaddim, u Speċifikazzjonijiet
Gwida komprensiva għall-Elitech MS-100 Digital Manifold Gauge, li tkopri l-karatteristiċi tiegħu, it-tħaddim rapidu, l-ispeċifikazzjonijiet tekniċi, il-prekawzjonijiet, u d-dettalji tal-prodott għas-sistemi tal-HVAC u tar-refriġerazzjoni.
Preview Manwal tal-Utent tal-Gauge tal-Manifold Diġitali tas-Serje MS ta' Elitech
Manwal tal-utent għall-gauge diġitali tal-manifold tas-Serje MS ta' Elitech, li jkopri d-deskrizzjoni tal-prodott, dettalji ġeneraliview, funzjonijiet tal-buttuni, speċifikazzjonijiet, gwida ta' bidu rapidu, passi ta' operazzjoni, applikazzjonijiet intelliġenti, u manutenzjoni.
Preview Manwal tal-Utent tal-Gauge tal-Manifold Diġitali Intelliġenti Elitech MS-100 PLUS
Manwal komprensiv għall-utent għall-Elitech MS-100 PLUS Smart Digital Manifold Gauge, li jkopri l-introduzzjoni tal-prodott, l-istruzzjonijiet tas-sigurtà, aktar minnview, funzjonijiet tal-buttuni, parametri tekniċi, u gwidi ta' tħaddim rapidu għal sistemi ta' refriġerazzjoni u HVAC.
Preview Elitech MS-100 Manwal għall-Utent tal-Gauge tal-Manifold Diġitali
Manwal tal-utent għall-Elitech MS-100 Digital Manifold Gauge, li jiddeskrivi l-karatteristiċi, l-operat, l-ispeċifikazzjonijiet u l-manutenzjoni tiegħu.
Preview Manwal tal-Utent u Speċifikazzjonijiet tal-Gauge tal-Manifold Diġitali tas-Serje Elitech MS-800
Manwal komprensiv għall-utent għall-istrument diġitali tal-manifold tas-Serje Elitech MS-800, li jkopri l-introduzzjoni tal-prodott, linji gwida dwar is-sigurtà, protezzjoni ambjentali, speċifikazzjonijiet dettaljati, u gwidi ta' tħaddim rapidu għall-kejl tal-pressjoni u t-temperatura, testijiet tal-pressjoni ta' żamma, u ċċarġjar tar-refriġerant għal sistemi HVAC u ta' refriġerazzjoni.
Preview Manifold Diġitali Elitech MS-100: Manwal tal-Utent u Speċifikazzjonijiet
Gwida komprensiva għall-manifold diġitali Elitech MS-100, li tkopri l-karatteristiċi tiegħu, l-ispeċifikazzjonijiet tekniċi, it-tħaddim, l-istruzzjonijiet tas-sigurtà, u s-soluzzjoni tal-problemi. Tgħallem kif tuża l-apparat għal kejl tal-pressjoni u t-temperatura, testijiet tat-tnixxija, proċessi tal-vakwu, u r-reġistrazzjoni tad-dejta bl-app Elitech Tools.