1. Introduzzjoni
This manual provides essential information for the safe and effective operation of the UEi IRT803 Digital Insulation Resistance Tester. The IRT803 is a versatile instrument designed for measuring insulation resistance, AC/DC voltage, and continuity. It features a 1000V AC/DC measurement capability, auto-ranging, a backlit display, and a DMM rotary style selector. Key functions include Lock, Compare, and Polarization Index (PI)/Dielectric Absorption Ratio (DAR) tests. Please read this manual thoroughly before use to ensure proper handling and to maximize the instrument's performance.
2. Informazzjoni dwar is-Sigurtà
Always adhere to safety precautions when using electrical testing equipment. The IRT803 is designed with safety in mind, featuring a CATIV rating up to 600V and CATII up to 1000V, ensuring protection in various jobsite environments. Failure to follow safety instructions can result in electric shock, fire, or damage to the instrument.
- Tużax it-tester jekk jidher li għandu l-ħsara jew jekk il-wajers tat-test huma kompromessi.
- Kun żgur li l-funzjoni u l-firxa korretti huma magħżula qabel ma tieħu xi kejl.
- Always disconnect power to the circuit under test before connecting or disconnecting test leads for resistance measurements.
- Ilbes tagħmir protettiv personali (PPE) xieraq, bħal ingwanti iżolati u nuċċalijiet tas-sigurtà.
- Do not operate the tester in wet conditions or in the presence of explosive gases or dust.
- Ibdel il-batteriji minnufih meta jidher l-indikatur ta' batterija baxxa biex tiżgura qari preċiż.
3. X'hemm fil-Kaxxa
Meta tiftaħ il-pakkett, ivverifika li l-oġġetti kollha elenkati hawn taħt huma preżenti u mhux imħassra:
- UEi IRT803 Insulation Resistance Tester
- Silicone test leads with alligator clips (red and black) (ATL57)
- Insulation test probe with alligator clips (red & black)
- 4 Batteriji AA
- Manwal għall-Utent
- Kaxxa tal-ġarr artab

Image 3.1: All components included with the UEi IRT803 Insulation Resistance Tester, including the meter, test leads, probes, batteries, manual, and soft carrying case.
4. Prodott Aktarview
The IRT803 is designed for ease of use and durability. It features an easy-to-read LCD display with a backlight for visibility in various lighting conditions, and a work light to illuminate dark environments. The unit is protected by a rugged rubber boot, which also includes convenient test lead holders for compact storage. An auto power-off function helps conserve battery life.
4.1. Karatteristiċi u Funzjonijiet Ewlenin
- 600V AC/DC Measurement: Capable of measuring both alternating and direct current voltaghuwa sa 600V.
- Reżistenza għall-iżolazzjoni: Measures resistance from 0.01MΩ up to 2GΩ with test voltages of 50V, 100V, 250V, 500V, and 1000V.
- Kontinwità: Għall-iċċekkjar tal-kontinwità taċ-ċirkwit.
- Compare Reading Function: Allows for comparison against a set reference value.
- PI/DAR Test: Automatic calculation of Polarization Index and Dielectric Absorption Ratio for advanced insulation diagnostics.
- A-Hold/Hold: Jiffriża l-qari muri għal reġistrazzjoni konvenjenti.
- Lock Reading: Locks the test voltage for continuous insulation testing.
- Rotary Dial Selector: Intuitive selection of measurement functions and test voltages.
- Stivali Protettivi tal-Lastku: Enhances durability and provides lead storage.

Stampa 4.1: Quddiem view of the IRT803 showing the LCD display, rotary selector, and function buttons.

Image 4.2: Visual representation highlighting the main functions and features of the IRT803.
5. Setup
5.1. Installazzjoni tal-batterija
The IRT803 requires four (4) AA batteries for operation. To install or replace batteries:
- Ensure the tester is powered off.
- Sib il-kompartiment tal-batterija fuq wara tal-unità.
- Uża tornavit biex tiftaħ l-għatu tal-kompartiment tal-batterija.
- Insert the four AA batteries, observing the correct polarity (+/-) as indicated inside the compartment.
- Agħlaq sewwa l-għatu tal-kompartiment tal-batterija.
5.2. Konnessjoni tal-wajers tat-test
Connect the test leads to the appropriate input jacks on the tester. For most measurements, the red lead connects to the 'V Ω INSULATION' jack and the black lead connects to the 'COM' (common) jack. Ensure connections are firm and secure.

Image 5.1: The IRT803 with test leads properly connected to the input jacks.
6. Istruzzjonijiet Operattivi
6.1. Powering On/Off and Display Backlight
- To power on the unit, rotate the selector dial from 'OFF' to any desired measurement function.
- To power off, rotate the selector dial back to 'OFF'. The unit also features auto power-off to save battery life after a period of inactivity.
- Agħfas il- AGĦŻEL button (often indicated by a light bulb icon) to toggle the display backlight on or off.
6.2. Ittestjar tar-Reżistenza għall-Insulazzjoni
The IRT803 can perform insulation resistance tests at various voltages.
- Select Test Voltage: Rotate the selector dial to the desired insulation test voltage (50V, 100V, 250V, 500V, jew 1000V).
- Qabbad il-wajers tat-test: Connect the red test lead to the circuit under test and the black test lead to the ground or common point. Ensure the circuit is de-energized.
- Ibda t-Test: Press and hold the large red TEST button to begin the insulation test. The test voltage will be applied, and the resistance reading will be displayed.
- Funzjoni Lock: For continuous testing, press the LOCK button after initiating the test. This will maintain the test voltage without needing to hold the TEST buttuna. Agħfas TEST għal darb'oħra biex tiftaħ.
- Compare Function: Uża l- Ipparaguna button to set a reference value and compare subsequent readings against it.
- PI/DAR Test: Select the PI/DAR function on the dial. The meter will automatically perform the Polarization Index and Dielectric Absorption Ratio tests, displaying the calculated values.
- Read and Record: Observe the reading on the LCD. Use the ŻOMM button to freeze the display for easier recording.
- Ħruġ: After completing an insulation test, the circuit under test should be discharged. The IRT803 typically handles this automatically, but always exercise caution.

Image 6.1: The IRT803 being used to measure insulation resistance on electrical wiring.

Immaġni 6.2: Close-up view of the test leads securely connected to the IRT803's input terminals.
6.3. Titjiratage Testing (AC/DC)
Biex tkejjel AC jew DC voltage up to 600V:
- Rotate the selector dial to the 'V' position.
- Connect the red test lead to the positive or live point and the black test lead to the negative or neutral/ground point of the circuit.
- Id-displej se juri l-volum imkejjeltage. The meter will automatically detect AC or DC voltage.
6.4. Ittestjar ta' Kontinwità
To check for continuity in a circuit:
- Rotate the selector dial to the 'Ω' (Ohms) position, which typically includes continuity.
- Connect the test leads across the component or circuit path you wish to test.
- If there is continuity (low resistance), the meter will typically emit an audible beep and display a low resistance value.
7. Manutenzjoni
7.1. Tindif
To maintain the accuracy and longevity of your IRT803:
- Imsaħ iċ-ċasing b'reklamaramp drapp u deterġent ħafif. Tużax abrażivi jew solventi.
- Kun żgur li l-unità tkun niexfa għalkollox qabel il-ħażna jew l-użu li jmiss.
7.2. Sostituzzjoni tal-batterija
Replace batteries as described in Section 5.1 when the low battery indicator appears on the display. Always use fresh AA alkaline batteries.
7.3. Ħażna
When not in use, store the IRT803 in its soft carrying case in a cool, dry environment, away from direct sunlight and extreme temperatures. If storing for extended periods, remove the batteries to prevent leakage.
8 Issolvi l-problemi
If you encounter issues with your IRT803, refer to the following common problems and solutions:
| Problema | Kawża Possibbli | Soluzzjoni |
|---|---|---|
| Tester does not power on | Batteriji mejta jew installati ħażin | Iċċekkja l-polarità tal-batterija; ibdel b'batteriji AA ġodda. |
| No reading on display | Incorrect function selected; open circuit; poor test lead connection | Verify selector dial position; ensure test leads are firmly connected to the circuit and the meter; check for breaks in the circuit. |
| Qari mhux eżatt | Low battery; environmental interference; incorrect range/function | Replace batteries; move away from strong electromagnetic fields; ensure correct function and test voltage are selected. |
| Test leads not working | Damaged leads; poor connection | Inspect leads for damage; ensure they are fully inserted into the jacks. Replace if damaged. |
If the problem persists after attempting these solutions, please contact UEi Test Instruments customer support.
9. Speċifikazzjonijiet
| Parametru | Valur |
|---|---|
| Numru tal-Mudell | IRT803 |
| Brand | Strumenti tat-Test UEi |
| Test tal-Insulazzjoni Voltages | 50V, 100V, 250V, 500V, 1000V |
| Firxa ta 'Reżistenza ta' Insulazzjoni | 0.01MΩ to 2GΩ |
| Insulation Resistance Accuracy | ±1.5-3% + 5 digits (0.01MΩ-500MΩ) |
| Voltage Kejl | Sa 600V AC/DC |
| Firxa ta' Kontinwità | 20.00 Ω to 20.0 kΩ (as per product description) |
| Klassifikazzjoni tas-Sigurtà | CATIV 600V, CATII 1000V |
| Sors tal-Enerġija | 4 x Batteriji AA |
| Wiri | LCD b'lura ta 'wara |
| Dimensjonijiet tal-Prodott | 2.7 x 3.52 x 8.27 pulzieri |
| Piż tal-oġġett | 1.8 Liri |
| Kulur | Isfar |
10. Informazzjoni dwar Garanzija
The UEi IRT803 Insulation Resistance Tester is backed by a Garanzija limitata ta 'sena from the date of purchase. This warranty covers defects in materials and workmanship under normal use. It does not cover damage resulting from misuse, unauthorized modification, accident, or neglect. For specific terms and conditions, please refer to the warranty card included with your product or contact UEi Test Instruments customer support.

Immaġni 10.1: Eż.ample of a Certificate of Conformity, indicating adherence to quality standards.
11. Appoġġ
For technical assistance, troubleshooting beyond this manual, or warranty claims, please contact UEi Test Instruments customer support. You can find more information and contact details by visiting the official UEi Test Instruments webis-sit jew il-paġna tal-ħanut tal-Amazon tagħhom.
UEi Test Instruments Store: Żur il-Ħanut