Apera Instruments AI531

Apera Instruments PH820 Benchtop pH Meter Kit

Manwal ta' Istruzzjoni

Prodott Overview

The Apera Instruments PH820 Benchtop pH Meter Kit is designed for reliable and accurate pH measurements in laboratory settings. It features advanced functionalities including intelligent calibration, data management, and a durable, water-resistant structure. This manual provides detailed instructions for proper use and care of your PH820 meter.

Apera Instruments PH820 Benchtop pH Meter Kit with electrode and stand

Figure 1: Apera Instruments PH820 Benchtop pH Meter Kit, showing the main unit, electrode, temperature probe, and electrode arm.

Key features of the PH820 include:

Setup

Spakkjar u Kontroll tal-Komponent

Carefully unpack all components from the kit. Verify that the following items are present:

Komponenti li jgħaqqdu

Connect the electrode, temperature probe, and power adapter to the corresponding ports on the back of the PH820 meter unit. Ensure all connections are secure.

Back panel of Apera Instruments PH820 pH meter showing connection ports

Figura 2: Wara view of the PH820 meter, highlighting the REF, pH/mV, TEMP, RS232, and DC 9V input ports.

  1. Konnessjoni tal-elettrodu: Plug the LabSen 211 pH Electrode into the port labeled "pH/mV".
  2. Konnessjoni tas-Sonda tat-Temperatura: Insert the temperature probe into the port labeled "TEMP".
  3. Konnessjoni tal-Enerġija: Connect the power adapter to the "DC 9V" port and then to a power outlet.
  4. Electrode Arm Assembly: Attach the flexible electrode arm and holder to the side of the meter unit for convenient electrode placement during measurements.

Istruzzjonijiet Operattivi

Basic pH Measurement

  1. Power on the meter by pressing the power button.
  2. Rinse the electrode with distilled water and gently blot dry.
  3. Immerse the electrode and temperature probe into the sample solution. Ensure the electrode tip is fully submerged.
  4. Wait for the reading on the large LCD to stabilize. The meter displays pH and temperature simultaneously.
  5. Record the stable reading.

Kalibrazzjoni tal-pH

The PH820 supports 1 to 5 points automatic calibration. Regular calibration ensures measurement accuracy.

  1. Prepare the standard calibration solutions (4.00 pH, 7.00 pH, 10.01 pH are included).
  2. Agħfas il- CAL buttuna biex tidħol fil-modalità tal-kalibrazzjoni.
  3. Rinse the electrode and immerse it into the first calibration solution (e.g., 7.00 pH).
  4. The meter will automatically recognize the solution and begin calibration. Wait for the stability icon to appear.
  5. Once stable, the meter will prompt for the next calibration point. Repeat the rinsing and immersion process for subsequent solutions.
  6. The self-diagnosis function will provide feedback on the calibration process and electrode condition, including slope data.
  7. Exit calibration mode once all desired points are completed.

Ġestjoni tad-Data

The PH820 allows for GLP data management, storing up to 500 groups of data. Data can be output via USB to a PC using the provided PC-Link software (Windows-based system only). Auto-timing data logging can also be configured.

ORP Measurement (Optional)

The PH820 can measure Oxidation-Reduction Potential (ORP) when an ORP electrode (e.g., 301Pt-C, sold separately) is connected to the pH/mV port. Refer to the ORP electrode's specific instructions for calibration and measurement procedures.

Manutenzjoni

Kura tal-elettrodu

Proper care of the LabSen 211 pH Electrode is crucial for maintaining accuracy and extending its lifespan.

Close-up of the LabSen 211 pH electrode

Figura 3: Close-up view of the LabSen 211 pH Electrode, showing its glass body and sensor tip.

Kura tal-meter

The PH820 meter unit is IP54 water-resistant, meaning it is protected from dust and splashing water. However, it should not be submerged. Keep the meter clean and dry. Use a soft, damp cloth to wipe the surface. Avoid using abrasive cleaners or solvents.

Issolvi l-problemi

The PH820 features a self-diagnosis function that can assist in identifying and fixing common problems, particularly during calibration. Refer to the meter's display for diagnostic messages.

ProblemaKawża PossibbliSoluzzjoni
Unstable ReadingsDirty electrode; air bubbles; insufficient sample volume; expired calibration.Clean electrode; ensure no bubbles; use adequate sample; recalibrate.
Nuqqas ta' KalibrazzjoniIncorrect buffer solution; contaminated electrode; electrode aging.Use fresh, correct buffers; clean electrode; check electrode slope data (self-diagnosis) for replacement indication.
Ebda Wiri/QawwaPower adapter issue; loose connection.Check power adapter and outlet; ensure all cables are securely connected.
Qari mhux eżattElectrode not calibrated; temperature not compensated; damaged electrode.Perform calibration; ensure temperature probe is functioning; inspect electrode for damage.

For persistent issues not resolved by the above, please contact Apera Instruments customer support.

Speċifikazzjonijiet

ParametruValur
Numru tal-MudellAI531
Eżattezza tal-pH±0.002 pH±1 digit
Punti ta' Kalibrar1 to 5 points (Auto.)
Ħażna tad-Data500 groups (GLP compliant)
Output tad-DataUSB (PC software support)
Reżistenza għall-IlmaIP54
Input tal-Enerġija120V AC 60Hz
Output ta 'Qawwa9V 300mA
Dimensjonijiet tal-Pakkett10.67 x 10.2 x 6.65 pulzieri
Piż6.69 Liri
ĊertifikazzjonijietCE, RoHS, ISO 9001:2015

Garanzija u Appoġġ

Apera Instruments products are manufactured under an ISO 9001:2015 certified quality system, ensuring high standards of production. For specific warranty information and technical support, please refer to the warranty card included with your product or visit the official Apera Instruments website. You may also contact their customer service directly for assistance with product operation, troubleshooting, or parts replacement.

Dokumenti Relatati - AI531

Preview Apera PH60Z-WW Smart pH Meter User Manual
Comprehensive user manual for the Apera PH60Z-WW Smart pH Meter, covering setup, calibration, measurement, troubleshooting, and specifications for wastewater, suspensions, and emulsions.
Preview Manwal tal-Utent tal-Meter tal-pH Intelliġenti PH60Z-HT - Apera Instruments
Manwal tal-utent għall-Apera Instruments PH60Z-HT Smart pH Meter, li jiddeskrivi l-karatteristiċi, it-tħaddim, il-kalibrazzjoni, il-manutenzjoni, u s-soluzzjoni tal-problemi għal soluzzjonijiet ta' temperatura għolja u kawstiċi.
Preview Apera SX811-SS Metru tal-pH Portabbli għall-Ittestjar tal-Ikel: Manwal tal-Istruzzjonijiet u Speċifikazzjonijiet
Manwal ta' struzzjonijiet komprensiv u speċifikazzjonijiet tekniċi għall-pH meter portabbli Apera SX811-SS, iddisinjat għal ittestjar preċiż tal-ikel. Tgħallem dwar il-karatteristiċi, il-kalibrazzjoni, is-setup, il-komunikazzjoni USB, u s-soluzzjoni tal-problemi.
Preview Apera PH60Z-MS Smart pH-Meter: Bedienungsanleitung für Kleinvolumenmessungen
Diese Bedienungsanleitung bietet detaillierte Informationen zur Verwendung des Apera PH60Z-MS Smart pH-Meters, einschließlich Kalibrierung, Messung, Wartung und Fehlerbehebung für präzise pH-Messungen in kleinen Volumina.
Preview Manwal tal-Utent tal-Meter tal-pH Apera PH60-HT: Soluzzjonijiet Kawstiċi u għal Temperatura Għolja
Manwal komprensiv għall-utent għall-Apera PH60-HT pH Meter, iddisinjat għal kejl preċiż tal-pH f'temperatura għolja (sa 140°C) u soluzzjonijiet kawstiċi. Ikopri l-kontenut tal-kit, it-tħaddim, il-kalibrazzjoni, il-kura tal-elettrodi, is-settings tal-parametri, l-ispeċifikazzjonijiet tekniċi, is-soluzzjoni tal-problemi, u l-garanzija.
Preview Apera PH700 Benchtop pH Meter Instruction Manual
Comprehensive instruction manual for the Apera PH700 Benchtop pH Meter, covering features, specifications, operation, calibration, maintenance, and troubleshooting.