Apera Instruments AI532

Apera Instruments EC820 Benchtop Conductivity Meter Kit

Manwal għall-Utent

1. Introduzzjoni u Prodott Fuqview

The Apera Instruments EC820 Benchtop Conductivity Meter is a high-precision instrument designed for accurate measurement of Conductivity (EC), Total Dissolved Solids (TDS), Salinity, Resistivity, and temperature. This meter is equipped with advanced features for reliable and stable readings across a wide range of applications.

This manual provides detailed instructions for the proper setup, operation, maintenance, and troubleshooting of your EC820 meter to ensure optimal performance and longevity.

Apera Instruments EC820 Benchtop Conductivity Meter Kit with electrode and beaker

Figure 1: Apera Instruments EC820 Benchtop Conductivity Meter Kit. The image shows the main meter unit, a flexible electrode arm and holder, and the conductivity electrode submerged in a beaker of liquid.

2. Prekawzjonijiet tas-Sigurtà

  • Always handle the meter and electrodes with care to prevent damage.
  • Do not immerse the main meter unit in water. The unit is IP54 water-resistant, but not submersible.
  • Uża biss l-adapter tal-enerġija pprovdut.
  • Keep calibration solutions and samples away from children and pets. Follow safety data sheets (SDS) for handling specific chemicals.
  • Ensure the meter is placed on a stable, level surface during operation.
  • Skonnettja l-enerġija qabel ma tnaddaf jew tagħmel manutenzjoni.

3. Kontenut tal-Pakkett

Iċċekkja li l-oġġetti kollha elenkati hawn taħt huma preżenti fil-pakkett tiegħek:

  • EC820 Preċiżjoni Benchtop Konduttività Meter
  • 2401T-F Platinum Glass-body Conductivity Electrode
  • Flexible Electrode Arm and Holder
  • Cable USB
  • PC-Link Software Flash Drive
  • Power Adaptor (Input: 120V AC 60Hz; Output: 9V 300mA)
  • Calibration Solutions:
    • 84 µS/cm (1 vial)
    • 1413 µS/cm (1 vial)
    • 12.88 mS/cm (1 vial)
Apera Instruments 2401T-F Platinum Glass-body Conductivity Electrode with cables

Figure 2: The 2401T-F Platinum Glass-body Conductivity Electrode, showing its connection cables.

4. Setup

  1. Unpack the Meter: Neħħi bir-reqqa l-komponenti kollha mill-imballaġġ.
  2. Assemble Electrode Arm: Attach the flexible electrode arm to the meter unit. Ensure it is securely fastened.
  3. Connect Electrode: Plug the 2401T-F Conductivity Electrode into the designated "COND" port on the back of the meter. Ensure the connection is firm.
    Lura view of Apera Instruments EC820 meter showing connection ports

    Figure 3: Rear panel of the EC820 meter, highlighting the "COND" and "TEMP" ports, along with RS232 and DC 9V power input.

  4. Qabbad l-Enerġija: Plug the power adapter into the "DC 9V" port on the back of the meter, then connect it to a standard 120V AC 60Hz power outlet.
  5. Qawwa Inizjali Mixgħula: Press the power button to turn on the meter. The LCD display should illuminate.

5. Istruzzjonijiet Operattivi

5.1. Kalibrazzjoni

The EC820 meter features 1 to 4 points auto-calibration. Regular calibration ensures accurate measurements.

  1. Prepare Solutions: Use the provided 84 µS/cm, 1413 µS/cm, and 12.88 mS/cm calibration solutions. Ensure solutions are at room temperature.
  2. Laħlaħ l-Elettrodu: Rinse the conductivity electrode thoroughly with distilled or deionized water before and after each measurement or calibration. Gently blot dry with a lint-free tissue.
  3. Daħħal il-Modalità tal-Kalibrazzjoni: With the meter on, press the CAL button. The display will show "CAL" and prompt for the first calibration point.
  4. First Point Calibration: Immerse the electrode into the 1413 µS/cm standard solution. Wait for the reading to stabilize (indicated by a stable icon on the display). The meter will automatically recognize the solution. Press DĦUL biex tikkonferma.
  5. Subsequent Points (Optional): For higher accuracy, proceed with 84 µS/cm and 12.88 mS/cm solutions following the same procedure. The meter will guide you through the process.
  6. Exit Calibration: After completing desired calibration points, press the MEAS button to return to measurement mode.

Nota: The meter has a self-diagnosis function to help identify and fix common calibration problems. Refer to the on-screen prompts.

5.2. Kejl

  1. Laħlaħ l-Elettrodu: Laħlaħ l-elettrodu b'ilma distillat.
  2. Immerse Electrode: Għaddas l-elettrodu fl-sampis-soluzzjoni, waqt li tiżgura li s-sensor ikun mgħaddas għalkollox.
  3. Wait for Stability: Allow the reading to stabilize. The stable reading icon will appear on the LCD.
  4. Qari tar-Rekord: Read the Conductivity (EC), TDS, Salinity, or Resistivity value displayed on the large LCD. Temperature will also be displayed simultaneously.
  5. Ibdel l-Unitajiet: Agħfas il- UNITÀ button to cycle through different measurement units (e.g., µS/cm, mS/cm, ppm, ppt).
  6. Ħażna tad-Data: The meter supports GLP data management with 500 groups of data storage. Use the M+ / R-M buttons to store and recall data.
  7. Esportazzjoni tad-Dejta: Connect the meter to a PC via the USB cable and use the provided PC-Link software (Windows-based system only) to export data.

6. Manutenzjoni

  • Tindif tal-Elettrodu:
    • Regularly rinse the electrode with distilled or deionized water.
    • For stubborn deposits, refer to the electrode's specific cleaning instructions. Generally, a mild detergent solution or a specific electrode cleaning solution can be used, followed by thorough rinsing.
    • Do not scratch or damage the platinum black sensor.
  • Ħażna ta 'elettrodi:
    • Always store the conductivity electrode in its protective cap filled with a proper electrode storage solution (e.g., 3M KCl solution or Apera Instruments storage solution). Never store dry.
  • Tindif tal-Meter: Wipe the main meter unit with a clean, damp drapp. Tużax tindif jew solventi li joborxu.
  • Aġġornamenti tas-Software: Check the Apera Instruments website periodically for any software updates for the PC-Link software.
Close-up of the Apera Instruments conductivity electrode tip

Figura 4: Close-up view of the platinum glass-body conductivity electrode tip, which requires careful handling and proper storage.

7 Issolvi l-problemi

ProblemaKawża PossibbliSoluzzjoni
Il-miter ma jixgħelxEbda provvista tal-enerġija; adapter difettuż; konnessjoni maħlula.Check power adapter connection and outlet. Ensure adapter is functional.
Qari instabbliDirty electrode; air bubbles on sensor; incorrect calibration; temperature fluctuations.Clean electrode; gently tap electrode to remove bubbles; recalibrate meter; ensure sample temperature is stable.
Żball fil-kalibrazzjoniContaminated calibration solution; incorrect solution used; damaged electrode.Use fresh calibration solutions; ensure correct solution for calibration point; inspect electrode for damage.
No data transfer to PCIncorrect USB connection; software not installed or running; driver issues.Verify USB cable connection; ensure PC-Link software is installed and open; check device manager for driver issues.

8. Speċifikazzjonijiet

ParametruValur
Numru tal-MudellAI532
Parametri tal-KejlConductivity (EC), TDS, Salinity, Resistivity, Temperature
Medda ta 'konduttività0 to 2,000 mS
Eżattezza± 0.5% FS
Punti ta' Kalibrar1 to 4 points auto-calibration (84µS/cm, 1413µS/cm, 12.88mS/cm, 111.8mS/cm)
Kumpens tat-TemperaturaKumpens Awtomatiku tat-Temperatura (ATC)
Ħażna tad-Data500 groups (GLP compliant)
KonnettivitàUSB for PC data export
Tip ta 'WiriLCD Backlit Kbir
Provvista tal-Enerġija9V DC, 300mA (120V AC 60Hz adapter included)
Protezzjoni tad-dħulIP54 Reżistenti għall-ilma
ĊertifikazzjonijietCE, RoHS, ISO 9001:2015
Dimensjonijiet tal-Pakkett10.63 x 9.88 x 6.61 pulzieri
Piż6.5 Liri

9. Garanzija u Appoġġ

Apera Instruments products are manufactured under an ISO 9001:2015 certified quality system, ensuring high standards of quality and reliability. For specific warranty information and technical support, please refer to the warranty card included with your product or visit the official Apera Instruments websit.

For assistance with troubleshooting, calibration, or general inquiries, please contact Apera Instruments customer service. Contact details can typically be found on their official websit jew fl-ippakkjar tal-prodott.

Riżorsi onlajn: Apera Instruments Store on Amazon

Dokumenti Relatati - AI532

Preview Apera EC20 Pocket Conductivity Tester: Manwal ta' Istruzzjonijiet u Gwida
Manwal ta' struzzjonijiet komprensiv għall-Apera EC20 Pocket Conductivity Tester. Tgħallem dwar is-setup, il-kalibrazzjoni, il-kejl tal-konduttività, l-ispeċifikazzjonijiet tekniċi, u l-informazzjoni dwar il-garanzija.
Preview APERA INSTRUMENTS PC950 Benchtop pH/Conductivity Meter Instruction Manual
Comprehensive instruction manual for the APERA INSTRUMENTS PC950 Benchtop pH/Conductivity Meter, covering specifications, operation, calibration, maintenance, and troubleshooting for laboratory use.
Preview Apera Instruments EC400S Portable Conductivity Meter Instruction Manual
Comprehensive instruction manual for the Apera Instruments EC400S Portable Conductivity Meter, covering features, specifications, operation, calibration, sample measurement, maintenance, parameter settings, and USB data communication.
Preview Apera Instruments SX731 Portable pH/Conductivity/ORP Meter User Manual
User manual for the Apera Instruments SX731 Portable pH/Conductivity/ORP Meter, covering its features, technical specifications, operation, calibration, maintenance, and warranty.
Preview Apera SX823-B pH/Conductivity Portable Meter Instruction Manual
This instruction manual provides detailed information on the Apera SX823-B portable pH/Conductivity meter, covering its features, specifications, operation, calibration procedures, maintenance, and troubleshooting.
Preview Manwal tal-Istruzzjonijiet tal-Apparat Instruments EC20 Pocket Conductivity Tester
Manwal tal-utent għall-Apera Instruments EC20 Pocket Conductivity Tester, li jkopri l-installazzjoni tal-batterija, il-funzjonijiet tat-tastiera, il-kalibrazzjoni, il-kejl tal-konduttività, l-ispeċifikazzjonijiet tekniċi, u l-informazzjoni dwar il-garanzija.