1. Introduzzjoni u Prodott Fuqview
The Apera Instruments EC820 Benchtop Conductivity Meter is a high-precision instrument designed for accurate measurement of Conductivity (EC), Total Dissolved Solids (TDS), Salinity, Resistivity, and temperature. This meter is equipped with advanced features for reliable and stable readings across a wide range of applications.
This manual provides detailed instructions for the proper setup, operation, maintenance, and troubleshooting of your EC820 meter to ensure optimal performance and longevity.

Figure 1: Apera Instruments EC820 Benchtop Conductivity Meter Kit. The image shows the main meter unit, a flexible electrode arm and holder, and the conductivity electrode submerged in a beaker of liquid.
2. Prekawzjonijiet tas-Sigurtà
- Always handle the meter and electrodes with care to prevent damage.
- Do not immerse the main meter unit in water. The unit is IP54 water-resistant, but not submersible.
- Uża biss l-adapter tal-enerġija pprovdut.
- Keep calibration solutions and samples away from children and pets. Follow safety data sheets (SDS) for handling specific chemicals.
- Ensure the meter is placed on a stable, level surface during operation.
- Skonnettja l-enerġija qabel ma tnaddaf jew tagħmel manutenzjoni.
3. Kontenut tal-Pakkett
Iċċekkja li l-oġġetti kollha elenkati hawn taħt huma preżenti fil-pakkett tiegħek:
- EC820 Preċiżjoni Benchtop Konduttività Meter
- 2401T-F Platinum Glass-body Conductivity Electrode
- Flexible Electrode Arm and Holder
- Cable USB
- PC-Link Software Flash Drive
- Power Adaptor (Input: 120V AC 60Hz; Output: 9V 300mA)
- Calibration Solutions:
- 84 µS/cm (1 vial)
- 1413 µS/cm (1 vial)
- 12.88 mS/cm (1 vial)

Figure 2: The 2401T-F Platinum Glass-body Conductivity Electrode, showing its connection cables.
4. Setup
- Unpack the Meter: Neħħi bir-reqqa l-komponenti kollha mill-imballaġġ.
- Assemble Electrode Arm: Attach the flexible electrode arm to the meter unit. Ensure it is securely fastened.
- Connect Electrode: Plug the 2401T-F Conductivity Electrode into the designated "COND" port on the back of the meter. Ensure the connection is firm.

Figure 3: Rear panel of the EC820 meter, highlighting the "COND" and "TEMP" ports, along with RS232 and DC 9V power input.
- Qabbad l-Enerġija: Plug the power adapter into the "DC 9V" port on the back of the meter, then connect it to a standard 120V AC 60Hz power outlet.
- Qawwa Inizjali Mixgħula: Press the power button to turn on the meter. The LCD display should illuminate.
5. Istruzzjonijiet Operattivi
5.1. Kalibrazzjoni
The EC820 meter features 1 to 4 points auto-calibration. Regular calibration ensures accurate measurements.
- Prepare Solutions: Use the provided 84 µS/cm, 1413 µS/cm, and 12.88 mS/cm calibration solutions. Ensure solutions are at room temperature.
- Laħlaħ l-Elettrodu: Rinse the conductivity electrode thoroughly with distilled or deionized water before and after each measurement or calibration. Gently blot dry with a lint-free tissue.
- Daħħal il-Modalità tal-Kalibrazzjoni: With the meter on, press the CAL button. The display will show "CAL" and prompt for the first calibration point.
- First Point Calibration: Immerse the electrode into the 1413 µS/cm standard solution. Wait for the reading to stabilize (indicated by a stable icon on the display). The meter will automatically recognize the solution. Press DĦUL biex tikkonferma.
- Subsequent Points (Optional): For higher accuracy, proceed with 84 µS/cm and 12.88 mS/cm solutions following the same procedure. The meter will guide you through the process.
- Exit Calibration: After completing desired calibration points, press the MEAS button to return to measurement mode.
Nota: The meter has a self-diagnosis function to help identify and fix common calibration problems. Refer to the on-screen prompts.
5.2. Kejl
- Laħlaħ l-Elettrodu: Laħlaħ l-elettrodu b'ilma distillat.
- Immerse Electrode: Għaddas l-elettrodu fl-sampis-soluzzjoni, waqt li tiżgura li s-sensor ikun mgħaddas għalkollox.
- Wait for Stability: Allow the reading to stabilize. The stable reading icon will appear on the LCD.
- Qari tar-Rekord: Read the Conductivity (EC), TDS, Salinity, or Resistivity value displayed on the large LCD. Temperature will also be displayed simultaneously.
- Ibdel l-Unitajiet: Agħfas il- UNITÀ button to cycle through different measurement units (e.g., µS/cm, mS/cm, ppm, ppt).
- Ħażna tad-Data: The meter supports GLP data management with 500 groups of data storage. Use the M+ / R-M buttons to store and recall data.
- Esportazzjoni tad-Dejta: Connect the meter to a PC via the USB cable and use the provided PC-Link software (Windows-based system only) to export data.
6. Manutenzjoni
- Tindif tal-Elettrodu:
- Regularly rinse the electrode with distilled or deionized water.
- For stubborn deposits, refer to the electrode's specific cleaning instructions. Generally, a mild detergent solution or a specific electrode cleaning solution can be used, followed by thorough rinsing.
- Do not scratch or damage the platinum black sensor.
- Ħażna ta 'elettrodi:
- Always store the conductivity electrode in its protective cap filled with a proper electrode storage solution (e.g., 3M KCl solution or Apera Instruments storage solution). Never store dry.
- Tindif tal-Meter: Wipe the main meter unit with a clean, damp drapp. Tużax tindif jew solventi li joborxu.
- Aġġornamenti tas-Software: Check the Apera Instruments website periodically for any software updates for the PC-Link software.

Figura 4: Close-up view of the platinum glass-body conductivity electrode tip, which requires careful handling and proper storage.
7 Issolvi l-problemi
| Problema | Kawża Possibbli | Soluzzjoni |
|---|---|---|
| Il-miter ma jixgħelx | Ebda provvista tal-enerġija; adapter difettuż; konnessjoni maħlula. | Check power adapter connection and outlet. Ensure adapter is functional. |
| Qari instabbli | Dirty electrode; air bubbles on sensor; incorrect calibration; temperature fluctuations. | Clean electrode; gently tap electrode to remove bubbles; recalibrate meter; ensure sample temperature is stable. |
| Żball fil-kalibrazzjoni | Contaminated calibration solution; incorrect solution used; damaged electrode. | Use fresh calibration solutions; ensure correct solution for calibration point; inspect electrode for damage. |
| No data transfer to PC | Incorrect USB connection; software not installed or running; driver issues. | Verify USB cable connection; ensure PC-Link software is installed and open; check device manager for driver issues. |
8. Speċifikazzjonijiet
| Parametru | Valur |
|---|---|
| Numru tal-Mudell | AI532 |
| Parametri tal-Kejl | Conductivity (EC), TDS, Salinity, Resistivity, Temperature |
| Medda ta 'konduttività | 0 to 2,000 mS |
| Eżattezza | ± 0.5% FS |
| Punti ta' Kalibrar | 1 to 4 points auto-calibration (84µS/cm, 1413µS/cm, 12.88mS/cm, 111.8mS/cm) |
| Kumpens tat-Temperatura | Kumpens Awtomatiku tat-Temperatura (ATC) |
| Ħażna tad-Data | 500 groups (GLP compliant) |
| Konnettività | USB for PC data export |
| Tip ta 'Wiri | LCD Backlit Kbir |
| Provvista tal-Enerġija | 9V DC, 300mA (120V AC 60Hz adapter included) |
| Protezzjoni tad-dħul | IP54 Reżistenti għall-ilma |
| Ċertifikazzjonijiet | CE, RoHS, ISO 9001:2015 |
| Dimensjonijiet tal-Pakkett | 10.63 x 9.88 x 6.61 pulzieri |
| Piż | 6.5 Liri |
9. Garanzija u Appoġġ
Apera Instruments products are manufactured under an ISO 9001:2015 certified quality system, ensuring high standards of quality and reliability. For specific warranty information and technical support, please refer to the warranty card included with your product or visit the official Apera Instruments websit.
For assistance with troubleshooting, calibration, or general inquiries, please contact Apera Instruments customer service. Contact details can typically be found on their official websit jew fl-ippakkjar tal-prodott.
Riżorsi onlajn: Apera Instruments Store on Amazon