Hantek DSO4102C/DSO4202C

Hantek DSO4102C/DSO4202C Digital Storage Oscilloscope User Manual

Models: DSO4102C (100MHz Bandwidth), DSO4202C (200MHz Bandwidth)

1. Introduzzjoni

The Hantek DSO4102C and DSO4202C are advanced dual-channel digital storage oscilloscopes that integrate an arbitrary/function waveform generator. These devices are designed for precise signal analysis and generation, featuring a large 7-inch 64K color LCD display with 800x480 resolution for clear and realistic waveform visualization. They offer a 1GSa/s sample rate and bandwidth options of 100MHz (DSO4102C) or 200MHz (DSO4202C). With over 20 types of automatic measurement functions and powerful trigger capabilities (Edge, Video, Pulse, Slope, Over time, Alternative), these oscilloscopes are versatile tools for various electrical and electronic applications. The integrated arbitrary/function waveform generator provides 25MHz output with 12-bit resolution and 200MHz DDS, capable of generating arbitrary, square, sine, triangle, trapezoidal, pulse, and DC waveforms to simulate transducers effectively. The devices support SD card storage and feature an integrated USB Host for USB disk storage and system updates.

Hantek DSO4102C/DSO4202C Oscilloscope and included accessories
Figure 1: Hantek DSO4102C/DSO4202C Oscilloscope with standard accessories including probes, USB cable, power cord, manual, and software CD.
Video 1: Hantek DSO4102C/DSO4202C Product Overview. This video demonstrates the physical appearance and some basic functionalities of the oscilloscope.

2. Setup

2.1 Spakkjar u Spezzjoni

Carefully unpack the oscilloscope and all accessories. Verify that all components listed in the packing list are present and undamaged. The standard package includes the oscilloscope unit, two oscilloscope probes, a USB cable, a power cord, a user manual, and a software CD.

Angolat view of Hantek DSO4202C Oscilloscope
Figura 2: Angolat view of the Hantek DSO4202C Digital Storage Oscilloscope.

2.2 Konnessjoni tal-Enerġija

  1. Ensure the oscilloscope is placed on a stable, flat surface with adequate ventilation.
  2. Connect the provided power cord to the AC power input on the rear panel of the oscilloscope.
  3. Plug the other end of the power cord into a suitable AC power outlet (AC 100-240V, 60W MAX).
  4. Press the power button on the front panel to turn on the device.
Wara view of Hantek oscilloscope showing power input and USB ports
Figure 3: Rear panel of the Hantek oscilloscope, highlighting the power input and USB ports. Note the 'Hazardous voltage inside' warning.

2.3 Konnessjoni tas-Sonda

  1. Connect the oscilloscope probes to the BNC input connectors on the front panel (CH1, CH2).
  2. Ensure the probes are securely twisted into place.
  3. Calibrate the probes before first use or if accuracy is critical. Refer to the operating section for probe calibration instructions.

2.4 USB and SD Card Connectivity

  • The oscilloscope features an integrated USB Host for connecting USB storage devices.
  • A USB interface is available for connecting to a computer for data transfer or system updates.
  • An SD card slot is provided for waveform storage and system updates.

3. Istruzzjonijiet Operattivi

3.1 Basic Oscilloscope Operation

The oscilloscope and waveform generator keys are separated for simultaneous operation. The large 7-inch 64K color LCD display (800x480 resolution) provides a clear view of waveforms.

Quddiem view of Hantek DSO4102C Oscilloscope
Figure 4: Front panel of the Hantek DSO4102C Digital Storage Oscilloscope, showing controls and display.

3.1.1 Vertical System Controls

  • Sensittività Vertikali: Adjusts the vertical scale (Volts/div) from 2mV/div to 10V/div.
  • Akkoppjar tad-Dħul: Selects DC, AC, or GND coupling for input signals.
  • Probe Attenuation: Set the probe attenuation coefficient (1X, 10X, 100X, 1000X) to match your probe.

3.1.2 Horizontal System Controls

  • Sampling Rate Range: Fixed at 1GSa/s for real-time samplipp.
  • Firxa tal-Iskennjar: Adjusts the horizontal time base (Sec/div) from 2ns/div to 80s/div (DSO4102C) or 4ns/div to 80s/div (DSO4202C).
  • Fond tal-Ħażna: 40K points.

3.1.3 Sistema ta' Trigger

The oscilloscope features powerful trigger functions to stabilize and capture specific events:

  • Trigger Types: Edge, Video, Pulse, Slope, Over time, Alternative.
  • Trigger Frequency Counter: Measures the frequency of the trigger source.
  • Miter tal-Frekwenza: Operates in "Pulse width" or "edge trigger" modes.

3.1.4 Funzjonijiet tal-Kejl

The device supports over 20 types of automatic measurement functions for various waveform parameters.

3.2 Arbitrary/Function Waveform Generator Operation

The integrated waveform generator can produce various signals:

  • Frekwenza: Sa 25MHz.
  • Riżoluzzjoni: 12-il bit.
  • Forom tal-mewġ: Arbitrary wave, square wave, sine wave, triangle wave, trapezoidal wave, pulse wave, DC.

4. Manutenzjoni

4.1 Kura Ġenerali

  • Keep the oscilloscope clean and free from dust. Use a soft, dry cloth for cleaning the exterior.
  • Evita li tesponi l-apparat għal temperaturi estremi, umdità, jew dawl tax-xemx dirett.
  • Żgura ventilazzjoni xierqa biex tevita sħana żejda.

4.2 Aġġornamenti tas-Software

The oscilloscope supports system updates via USB interface or SD card. Periodically check the Hantek official website for the latest firmware and software to ensure optimal performance and access to new features.

5 Issolvi l-problemi

If you encounter issues with your Hantek oscilloscope, consider the following basic troubleshooting steps:

  • Ebda Qawwa: Check the power cord connection and the power outlet. Ensure the power button is pressed.
  • L-ebda Wirja tal-Forma tal-Mewġa: Verify that probes are correctly connected to both the oscilloscope and the signal source. Check input coupling and vertical/horizontal scale settings. Ensure the trigger is set appropriately.
  • Forma tal-Mewġa Distorta: Check probe compensation. Ensure the input voltage does not exceed the maximum rating (Type I and Type II: 300VRMS (10x); Installation type III: 150VRMS (1x)).
  • Ħsara fis-sistema: Try restarting the device. If the issue persists, consider performing a factory reset (refer to the on-screen menu or detailed manual for instructions) or updating the firmware.
  • Karta USB/SD Mhux Rikonoxxuta: Ensure the storage device is properly inserted and formatted correctly (usually FAT32).

For more complex issues, consult the comprehensive user manual provided on the software CD or contact Hantek customer support.

6. Speċifikazzjonijiet

Below are the detailed specifications for the Hantek DSO4102C and DSO4202C models:

MudellDSO4202SDSO4102SDSO4072S
Oscilloscope Index
Sampling TipĦin reali sampling: 1GSa/s
Ekwivalenti sampling: 25GSa/s
Mod ta' Akkwist
SamplippSampdejta dwar il-linarda
Peak to PeakDisplay high frequency and random burrs
MedjaAverage waveform, times: 4, 8, 16, 32, 64, 128
Input
Akkoppjar tad-DħulDC, AC or ground (DC, AC, GND)
Impedenza tad-dħul1MΩ±2% / 120pF±3pF
Attenwazzjoni tas-Sonda1X, 10X
Issettjar tal-Koeffiċjent tal-Attenwazzjoni tas-Sonda1X, 10X, 100X, 1000X
Input VoltageType I and Type II: 300VRMS (10×); Installation type III: 150VRMS (1×);
Installation type II; With the slope of 20 dB/decade, the AC peak value is 13V when it starts to drop from above 100kHz to more than or equal to 3MHz *. For non sinusoidal waveform, the peak value must be less than 450V. The offset duration above 300 V shall be less than 100 ms, and the RMS signal level (including all DC components deleted through AC coupling) must be limited to 300 V. If these values are exceeded, the instrument may be damaged.
Sistema Orizzontali
SampLing Rata Medda1GS / s
Interpolazzjoni tal-forma tal-mewġ(sin x)/x
Fond tal-Ħażna40K
Firxa tal-Iskanjar2ns/div~80s/div4ns/div~80s/div
Sampling Rate and Delay Time Accuracy± 50ppm (at any time interval greater than 1ms)
Medda tal-Pożizzjoni2ns/div to 10ns/div; (-4div x s/div) to 20ms;20ns/div to 80us/div; (-8div x s/div) to 40ms; 200us/div to 40s/div; (-8div x s/div) to 400s
Time Interval Measurement Accuracy (full bandwidth)Ħin wieħed: ± (1 sampling interval+100ppm x reading+0.6ns); >16 average values: ± (1 sampling interval+100ppm × Reading+0.4 ns); Sampling interval=seconds/division ÷ 200
Sistema Vertikali
Riżoluzzjoni Vertikali8-bit resolution, simultaneous sampling for each channel
Sensittività Vertikali2mV/div to 10V/div
Bandwidth200MHz100MHz70MHz
Rise Time at BNC (typical)1.8ns3.5ns5ns
Analog Bandwidth in Sampling and Averaging Mode, DC Coupling at BNC or with Probe2mV/grid to 20mV/grid, ± 100mV; 50mV/grid to 200mV/grid, ± 1V; 500mV/grid to 2V/grid, ± 10V; 10V/grid, ± 50V
Operazzjoni MatematikaAdd, subtract, multiply, divide, FFT
FFTWindows: Hanning, Flat, Rectangular, Bartlett, Blackman; 1024 samppunti ling
Limitazzjoni tal-bandwidth20MHz
Low Frequency Response (-3db)≤ 10Hz at BNC
Vertical Gain AccuracyIn the "Sampling" or "Average" acquisition mode, the accuracy from 10V/grid to 10mV/grid is ± 3%; In the "Sampling" or "Average" acquisition mode, the accuracy from 5mV/grid to 2mV/grid is ± 4%
Trigger Frequency Counter
Riżoluzzjoni tal-Qari6 ċifra
Precision (typical)± 30ppm (including all frequency reference errors and ± 1 calculation error)
Medda ta' FrekwenziAC coupling, from 4Hz to rated bandwidth
Metru tal-Frekwenza"Pulse width" or "edge trigger" mode: all available trigger sources. The "frequency counter" has been measuring the trigger source, including when the oscilloscope acquisition is suspended due to changes in the operating state, or when the single event acquisition has ended. "Pulse width" trigger mode. The oscillograph calculates the pulse with effective amplitude in the window measured in 1s and meeting the trigger event conditions. For example, if the PWM pulse column is set to <mode and the width is set to a smaller time, the narrow pulse will appear. "Edge trigger" mode: the oscilloscope calculates all. High frequency suppression: signal will be attenuated when it is greater than 80kHz. Low frequency suppression: when the frequency is greater than 150kHZ, it is the same as the DC coupling limit, less than 150kHZ.

Proprjetajiet Addizzjonali tal-Prodott:

  • Batterija Inkluża: Nru
  • Kimika ta' tħassib kbir: Xejn
  • IS Battery Demountable: Nru
  • Tul tar-Rekord: 40K
  • Numru tal-Mudell: DSO4102C/DSO4202C
  • Ċertifikazzjoni: CE
  • Oriġini: Iċ-Ċina kontinentali
  • Ħin Real SampRata tal-linarda: 1GSa/S
  • Provvisti DIY: Elettriku
  • Kanali Diġitali: 2
  • HANTEK Model: DSO4102C/DSO4202C
  • Riżoluzzjoni tal-wiri: 800 * 480 Pixels
  • Daqs tal-Wiri: 7 Pulzieri u 'l fuq
  • Dimensjonijiet ġenerali: 385mm * 200mm * 245mm
  • Pakkett: Iva

7. Garanzija u Appoġġ

Hantek products typically come with a manufacturer's warranty. For specific warranty terms and conditions, please refer to the documentation included with your product or contact Hantek customer support directly. Support may include technical assistance, troubleshooting guidance, and repair services.

8. Għajnuniet għall-Utenti

Ma kien hemm l-ebda pariri speċifiċi għall-utent disponibbli minn reviews or Q&A for this product in the provided data. For best results, always ensure your probes are properly calibrated and matched to the oscilloscope's settings. Regularly check for firmware updates to enhance functionality and stability.

Dokumenti Relatati - DSO4102C/DSO4202C

Preview Manwal tal-Utent tal-Oxxilloskopju Diġitali tas-Serje Hantek DSO2D20
Manwal komprensiv għall-utent għall-oxxilloskopji diġitali tas-serje Hantek DSO2D20, li jiddeskrivi l-karatteristiċi, it-tħaddim, is-soluzzjoni tal-problemi, u l-ispeċifikazzjonijiet għal analiżi preċiża tas-sinjali.
Preview Manwal tal-Utent tal-Oxxilloskopju Dijanjostiku tal-Karozzi Hantek2D82 | Gwida Komprensiva
Esplora l-Oxxilloskopju Dijanjostiku tal-Karozzi Hantek2D82. Dan il-manwal tal-utent jipprovdi informazzjoni dettaljata dwar il-karatteristiċi, l-operat, is-sigurtà, l-ispeċifikazzjonijiet tekniċi tiegħu, u s-soluzzjoni tal-problemi għad-dijanjostika tal-karozzi, id-DMM, u l-ġenerazzjoni tal-forma tal-mewġ.
Preview Hantek DSO2D20 Series Digital Oscilloscope Quick Guide
A concise and SEO-optimized HTML guide for the Hantek DSO2D20 Series Digital Oscilloscope, covering safety, product overview, preparation, and remote control.
Preview Hantek DSO3062L/DSO3062AL Digital Oscilloscope User Manual
Explore the Hantek DSO3062L/DSO3062AL Digital Oscilloscope User Manual. This guide provides detailed instructions on setup, operation, safety precautions, advanced functions, and application examples for accurate electronic measurements.
Preview Hantek2000 Series Handheld Scopemeter User Manual V1.1
This user manual provides comprehensive instructions and specifications for the Hantek2000 Series Handheld Scopemeter, covering its features, operation, safety guidelines, troubleshooting, and technical details.
Preview Manwal tal-Utent Hantek 6022BE: Gwida għall-Oxxilloskopju tal-PC Portabbli
Niżżel il-manwal tal-utent Hantek 6022BE għal dan l-oxxilloskopju USB portabbli bbażat fuq il-PC. Tgħallem dwar il-karatteristiċi, is-setup, it-tħaddim, u s-soluzzjoni tal-problemi tiegħu għal applikazzjonijiet ta' ttestjar, riċerka, u edukazzjoni.